Since the color of the surface may be faded in repetitive
processing, do not use products that cannot be clearly
distinguished with color coding.
3) Packaging
Place clip in the designated clip case or tray and close the
lid.
Then pack the case (tray) containing the product using
standard medical packaging material (non-woven fabric
or sterilized bag) with protective sterile barrier complying
with ISO 11607 for steam sterilization.
- Validated Sterilization Procedure -
With the pre-vacuum type high-pressure steam sterilizer
which functions properly, this product shall be sterilized
so that sterilization conditions [134 °C for 3 minutes or
more] are maintained.
Do not sterilize using low-temperature hydrogen
peroxide gas plasma. It may cause discoloration and/or
deformation.