Dantec Keypoint
• Make sure that only fuses with the required
rated current and of the specified type are
used for replacement. The use of makeshift
fuses and the short-circuiting of fuse holders
are prohibited.
• Where more than one piece of equipment is
connected to a patient, attention must be paid
to the summation of patient leakage currents.
• Whenever it is likely that the protection has
been impaired, the device shall be made
inoperative and be secured against any
unintended operation.
Call qualified service personnel to conduct at
least a functional test and additionally a safety
check including 1) an insulation test; 2) a
ground continuity test; and 3) a leakage
current test according to IEC 60601-1.
• The protection is likely to be impaired if, for
example, the device:
- Shows visible damage;
- Fails to perform the intended function(s);
- Has been subjected to severe transport
stress.

Connection Panel

Models: 9031D040x / 9031D041x
CAUTION In the USA: If 240V~, only center-
tapped, single phase circuit.
NOTE Unplugging the power line cable from
the mains input on the Isolating Transformer,
disconnects the mains power to all supplied
devices.
(1)
Mains Input
Plug the shielded mains power cord
supplied with the system into this socket.
NOTE Only use a shielded power cord from
Natus.
Rating Frequency: 50/60Hz.
- Continuous Operation: max. 1000VA
- Momentary Operation: max. 1500VA
9031D040x: 100/120V~
9031D041x: 200/240V~
(2)
Fuse
F1 and F2 Mains
9031D040x: 110/120V: T10A / 250V~
9031D041x: 200/240V: T6.3A / 250V~
(3)
M4 Thread for Electrical Enclosure
Contact and Additional Protective Earth.
Potential equalization
(4)
Where potential equalization is necessary or
desirable, a potential equalization conductor
may be connected to this connector as per
DIN 42 801.
(5)
Spring Retaining Clamp for Power Cord
Place the spring retaining clamp around the
power cord to ensure a safe connection.
(6)
Auxiliary Outlet – 6 Outlets: 50/60Hz
9031D040x:
100-120V~ total max.
- Continuous Operation: 7.4A
- Momentary Operation: 12.4A
7.4A equals 850VA, and 12.4A equals
1425VA at 115V~
9031D041x:
200-240V~ total max.
- Continuous Operation: 3.7A
- Momentary Operation: 6.2A
3.7A equals 850VA and 6.2A equals
1425VA at 230V~
(7)
SN: Device Serial Number
(8)
REF: Device Reference Number

Environmental Conditions

Operating Environment
Temperature:
+10°C to +35°C (+50°F to +95°F)
Relative Humidity:
20% to 80% (non-condensing)
Altitude:
-15m to 3,000m
Pressure:
701hPa to 1060 hPa
– 6 –