6. MOUNTING AND FITTING ON THE PATIENT
A. Assembly
Adhere to the tightening torque of the M10 screw: 22Nm
Adhere to the engaged screw thread length: 10 +/-1 mm
7. DETECTING MALFUNCTIONS
If you observe abnormal behaviour or feel modifications in the device's properties, or if it undergoes a heavy impact, please consult your
orthopaedic prosthetist.
8. WARNINGS, CONTRAINDICATIONS, SIDE EFFECTS
A. Warnings
Using the device in a way that goes against the advice given by your orthopaedic prosthetist may cause deterioration of its components
(e.g. bearing heavy loads, excessive strain, using beyond the service life, etc.).
The device is weatherproof only.
B. Contraindications
This device is not intended to be used for activities associated with a risk of significant impact or excessive loading.
This device is not resistant to fresh water, salt water or chlorinated water.
C. Side effects
There are no side effects directly linked to the device.
Any serious incident arising in relation to the device must be reported to the manufacturer and to the competent authority of the
Member State.
9. CARE, STORAGE, DISPOSAL AND SERVICE LIFE
A. Care/ cleaning
Clean the device with a damp cloth or sponge.
No greasing, adjustments to the fasteners or any other operation is required for this device, except by the orthopaedic prosthetist.
After accidental contact with water:
•
Rinse the device in clean water
•
Dry thoroughly
The front stop and rear shock absorber may require replacement.
1
B. Storage
Usage and storage temperature: -20°C to +60°C
Relative air humidity: no restrictions
C. Disposal
The various components of the device are special waste: Aluminium, steel, thermoplastic, polyurethane elastomer, natural rubber.
They must be treated in accordance with current, applicable legislation.
D. Service life
It is advisable to arrange for an orthopaedic prosthetist to perform an annual check.
USER MANUAL
2
3
Page 2 of 3
4