2
Instructions for use for Prosthetists/Orthotists
0
1
6
Please read before using or working on the device
It is important to read and explain the patient information leaflet to the patient.
1. COMPONENTS INCLUDED
Designation
Knee
Protective hood
2.5 mm Allen (hex) key
Cover
One-part cosmetic
cover
2. DESCRIPTION, PROPERTIES AND MECHANISM OF ACTION
A. Description
A single-axis knee offering hydraulic regulation of the stance and swing phases, fitted with a manual lock that halts
the flexion of the knee.
It is available in four different versions:
•
1P130/1P131: Male pyramid proximal connector, suitable for patients who have had a transfemoral amputation or hip
disarticulation
•
1P130-KD/1P131-KD: Threaded proximal connector (M36×1.5), suitable for patients with a long transfemoral amputation stump
or who have had a knee disarticulation
For all 4 versions, the distal connection is made via a Ø34 mm tube
B. Properties
Ref.
Colour
Weight
Maximum flexion
Maximum patient weight
(Including carried load)
This device has undergone a three-million-cycle test, in accordance with standard NF EN ISO 10328 at a P7 load level (i.e. 150 kg),
which corresponds to a service life of 4 to 5 years, depending on the patient's mobility level.
HYTREK – 1P130(-KD) / 1P131(-KD)
Ref.
1P130(-KD) or 1P131(-KD)
1P13055
WKF026
1P13029-B/-N
1P13028-B/-N
1P13027-B/-N
1G16
1P131
Black
36 mm
Ø34
1P130/1P131
Included/Sold separately
Included
Included
Included
Included
Sold separately
1P130
1P131-KD
Blue
Black
1265 g
1285 g
120°
150 kg
Axis of rotation
33 mm
1P130-KD/1P131-KD
1P13099
2021-07
1P130-KD
Blue
Ø34