8. GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS
The handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of
the handpiece should assure that it is used in such an environment.
Emissions test
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations
Not applicable
/flicker emissions
IEC 61000-3-3
RF emissions
Complies
CISPR 14-1
The handpiece is intended for use in the electromagnetic environment specified below. The Customer or the user of
the handpiece should assure that it is used in such an environment.
Immunity test
Electrostatic
±6 kV contact
discharge
±8 kV air
(ESD) IEC 61000-
4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1KV for input/
output lines
Surge
±1 kV differential
IEC 61000-4-5
mode
±2 kV common
mode
Voltage dips, short
<5 % UT for 0,5
interruptions and
cycle
voltage variations
(>95 % dip in UT )
on
power supply input
40 % UT for 5
lines
cycles
IEC 61000-4-11
(60 % dip in UT )
70 % UT or 25
cycles
(30 % dip in UT )
<5 % UT for 5 sec
(>95 % dip in UT )
The handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of
the handpiece should assure that it is used in such an environment.
ZF1902109.X / 03/ 2009 –updated 03/2011
Chemin du Verger 3
CH 1338 Ballaigues
Switzerland
www.dentsplymaillefer.com
Compliance
The handpiece uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The handpiece is not suitable for interconnection with other equipment.
IEC 60601
Compliance level
test level
±2, ±4, ±6KV
Contact
±2, ±4, ±8KV Air
Not applicable
Not applicable
Not applicable
Electromagnetic environment – guidance
Electromagnetic environment –guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered with
synthetic material, the relative humidity should
be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the handpiece requires continued operation
during power mains interruptions, it is
recommended that the handpiece be powered
from an uninterruptible power supply or a
battery.