SYSTEM DESCRIPTION

The OLM Intracranial Pressure Monitoring
Kit consists of a sterile transducer-tipped
pressure monitoring catheter with accessory
items to be used as a diagnostic tool for rapidly
determining
and
continuously
intracranial pressure. Since its method of
measuring pressure is unique, please read this
section carefully.
Unlike ordinary pressure monitoring systems,
the
Camino
Catheter
®
transducer at the distal tip. This unique design
eliminates the need for a "fluid-filled system" to
carry pressure waves to an external transducer.
Thus, the problems inherent in such fluid-filled
systems are eliminated. The result is a precise
pressure measurement and an artifact free,
high fidelity waveform trace.
The OLM Intracranial Pressure Monitoring Kit
contains the following accessory items for use
with the Camino Catheter (Figure 1);
• Camino Bolt with Compression Cap and
Turning Wings,
• 0.106 inches (2.71 mm) diameter No. 36 Drill
Bit with Safety Stop,
• Spacer to adjust seating depth of the bolt,
• Strain Relief — Protective Sheath,
OLM* INTRACRANIAL
PRESSURE MONITORING KIT
MODEL 110-4B
SAFETY STOP
SAFETY STOP
DRILL BIT
DRILL BIT
TRANSDUCER
TRANSDUCER
AIR VENT
AIR VENT
STRAIN RELIEF SHEATH
STRAIN RELIEF SHEATH
1 cm MARKINGS
1 cm MARKINGS
UP TO 10 cm
UP TO 10 cm
monitoring
has
a
miniature
ZERO ADJUSTMENT TOOL
ZERO ADJUSTMENT TOOL
• 0.050 inches (1.3 mm) diameter Stylet,
• Hex wrench,
• Zero adjustment tool
Note: A hand drill and various standard surgical
instruments and supplies are required to place
the ICP monitoring system.
INDICATIONS
The use of the OLM* Intracranial Pressure
Monitoring Kit by a qualified neurosurgeon
is indicated when direct measurement of
intracranial pressure in the parenchyma or the
subarachnoid space, is clinically important.

CONTRAINDICATIONS

• This device is not intended for any use other
than that indicated.
• Magnetic
Safety Information: The Camino 110-4B
is MR Unsafe. Do not bring catheter or
accessories into the MR environment.