NSK Surgic XT Plus Mode D'emploi page 13

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

(7) Speed calibration: "CAL Lmin
calibration bur is removed and the handpiece is free to rotate
(Fig. 20) ; Press the [SYSTEM] Key after a notification sound, the
calibration is automatically started, the handpiece will illuminate
(Optic Handpiece only) and start to rotate, it will automatically
cycle through to the highspeed test. On completion the motor
stops "donE" is displayed the calibration is complete, normal
operation mode will continue.
CAUTION
During the process, the rotational speed is automatically increased and finally rotates at high speed.
Significant care should be taken to avoid danger.
(8) Calibration of the handpiece is completed.
Calibration fail example
• Extreme gear abrasion and mechanical loss can exceed the control limits. Correct data cannot be obtained
during calibration cycle and the calibration will fail. In this case, request repair.
• Calibration fail; if the calibration bur makes contact with something during no load calibration or no load is
erroneously applied during load calibration.
6-3 Optical Illumination option On/Off (X-SG20L Only)
* Factory initialized settings lamp 'ON'
(1) Selection
1) Lamp OFF: Hold the [System] Key and turn the power 'ON'; Two beeps notifies the operator the lights are 'OFF'
2) Lamp ON: Press the [System] Key and turn the power 'ON'; One beeps notifies the operator the lights are 'ON'
(2) Confirmation
Confirm whether the lamp of the motor lights or not by the beep issued when turning the power ON.
– Long beep once: "Lights up"
– Short beep twice: "Does not light up"
NOTICE
• Repeat the above procedures when restoring operation.
• The lamp status will remain set even if the power is switched "OFF' and 'ON.
• This function is only applicable for an optic handpiece.
6-4 Standard Operation
All standard operational functions can be control at the Foot Control.
(1) Turn on the Main Power Switch: The Control Unit is ready to perform the Program.
(2) Select the desired program number: Step on the Foot Control PRG (Program) Button and the program display
ascends to the next program number. Pressing the PRG (Program) Button for one second more will descend the to
the next program number.
(3) Verify the details of the program on the display.
-1
" is displayed. Ensure the
12
The calibration bur is removed.
The calibration bur is removed.
The calibration bur is removed.
Fig. 20

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sgl50mSg 50msX-sg20lSg20Ne179E295-050 ... Afficher tout

Table des Matières