8. Troubleshooting Guide

The following table contains fault symptoms, their possible causes and suggests
steps to rectify the problem. These may occur during initial installation and power
up.
Symptom
Pump keyboard does
not respond.
Mattress does not
inflate when the
system is switched
on.
Mattress inflates but
appears to be noisy as
if a leak has appeared
in the tube-set
connection with the
pump.
Mattress inflates but
doesn't appear to turn.
Mattress inflates but
only appears to turn to
one side.
Mattress turns but
then suddenly stops
mid cycle.
Possible Cause
Lock out button LED
Lock out is on
illuminated.
preventing set
up.
If the pump is plugged into a
Pump is left in
power socket, a dash will be
standby mode,
visible on the LED display to
indicated by a
indicate standby mode.
single dash on
Check status of LED display.
the LED display.
Check to ensure blower is
Blower failure
generating airflow through
within pump unit.
the manifold connector.
Check to ensure the hose
Hose-set
connector clicks into place.
connection not
correctly fitted.
Check plug is fitted correctly
Bolster/Sub-
in hose manifold.
Mattress plug not
fitted correctly or
Check integrity of plug.
failing.
Therapy Settings.
No turn angle
set.
Firm inflate
Maxflow
activated.
Check status of LED display.
Mattress remains
central (check
turn time).
Ensure bilateral therapy
Therapy is set for
setting has been set.
left or right turn
only.
For bilateral turn ensure
Both setting may
not be
illuminated.
illuminated).
LED display showing '
Maxflow
activated.
Check status of LED display.
Mattress remains
central (check
turn time).
Check
LED illuminated.
is activated (LED
'.
F
15
Action
Press and hold the lock
button until the LED
extinguishes.
Turn on the pump unit.
If the LED display is blank,
check integrity of power
supply at the wall socket.
If blower has failed call a
Service Engineer to replace
the blower/pump unit.
Re-connect hose connection
to pump, ensuring firm
fitment.
Ensure plug is firmly located.
Replace if neccesary.
Set
angles.
Turn
Press the
button to
deactivate, and enter
required therapy settings.
If LED display is illuminated
with '
' mattress is
c
centralised.
Enter
therapy
Bi-lateral
settings.
If not, press
to select
and ensure the LED
illuminates.
Press the
button to
deactivate, and set turn
angles.
If LED display is illuminated
with '
' mattress is
c
centralised.