Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 2)
Electromagnetic immunity
The Symphony electric breastpump is intended for use in the electromagnetic environment
specifi ed below. The customer or the user of the Symphony electric breastpump should assure
that it is used in such an environment.
Immunity test
Electrostatic
discharge ( ESD )
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Questions? Visit www.medela.com or call us at 1-800-435-8316.
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 1 kV line( s ) to
line(s)
± 2 kV line( s ) to
earth
Electromagnetic
Compliance
environment
level
– guidance
Floors should be wood,
concrete or ceramic tile.
± 6 kV contact
If fl oors are covered with
± 8 kV air
synthetic material, the
relative humidity should
be at least 30%.
± 2 kV for power
Mains Power quality
supply lines
should be that of a
± 1 kV for input/
typical commercial or
output lines
hospital environment.
± 1 kV differential
Mains Power quality
mode
should be that of a
± 2 kV common
typical commercial or
mode
hospital environment.
39