9.3.7. Interval metering

Interval metering is a sub-routine of metering by volume mode; repetition is not launched by
-
hitting a key, or by a triggering pulse, but automatically after a pause selected before.
Set volume as described under 10.3.5, metering by volume.
-
Hold Select-key (36) depressed until „Pause sec/min/h" LED lights.
-
Enter pause with arrow keys (35); choose pause between 0.1 seconds and 750 hours.
-
seconds range: display 0.1 to 599.9 (no extra signal displayed)
minutes range: display 10 to 599, „' " displayed(600sec= 10min)
hours range: display 10 to 750 „h" displayed (600min=10Std.)
Hold select key (36) depressed until „Vol ml" LED and LED „Pause sec/min/h" LED light the
-
same time.
Interval metering is controlled with the „Start Stop" key (32) or via the interface.
-
After having hit the „Start Stop" key (32) the metered volume is displayed (41), rising from
-
zero to the current value, or a pause interval rising from zero the value s et before. The
respective LED is going to flash.
9.3.8. Adjust backstroke for metering by volume mode (figs. 49)
To avoid dripping in the "metering by volume" mode, a short backstroke of the pump may be
programmed.
Turn power switch (29) OFF. Hit „Cal" key (34), and, while holding depressed, turn power
-
switch (29) ON.
Pump head, e.g. "Pu 1" is displayed (41).
-
Hit arrow keys (35) until „r on" or „r of" is displayed (41).
-
„r on" turns backstroke ON, „r off" turns backstroke OFF.
-
Confirm mode with select key (36).
-
9.4. Interfaces (fig. 50)

9.4.1. Analog interface

An analog interface connector (37) is installed in the item's rear panel.
-
External pump control uses this analog interface.
-
Control voltage of 0 - 10 volts equals rotational speed between 0 and max.
control voltage
0 - 0.9 V
1 V
10 V
Control current 4 - 20 mA equals rotational speed between 0 and max.
control current
0 - 4.9 mA
5 mA
20 mA
PD 5201
0 [1/min]
5 [1/min]
120 [1/min]
PD 5201
0 [1/min]
5 [1/min]
120 [1/min]
65
PD 5206
0 [1/min]
24 [1/min]
606 [1/min]
PD 5206
0 [1/min]
24 [1/min]
606 [1/min]