Télécharger Imprimer la page

Struers RepliSet Mode D'emploi page 4

Publicité

Use
• After use leave the mixing nozzle in place on the cartridge to act
as a seal.
• Each time the fl ow of compound is stopped for a period longer
than the working life (see cartridge) a new nozzle must be used.
Removal of Cartridge
• Push slide release lever and pull the piston slide to its back
position. Lift the cartridge clip and remove the cartridge.
Cleaning of the Surface
The area to be examined should be clean and degreased. Use
cotton wool and a solvent.
• Alcohol will normally be suffi cient for metallographic applications.
• Engineering surfaces should be degreased using a suitable
solvent like acetone.
• Make sure that the solvent has evaporated completely before
application of the compound.
Note: RepliSet will not cure if residues of grease or solvents are
left on the surface or retained in inclusions.
Selection of Compound
Surface facing
Horizontal,
sloping
Vertical, overhead > 20°C
The process time varies with the temperature. Please see graph in
the RepliSet Instruction Manual.
Temperature of
surface
> 20°C
< 20°C
< 20°C
Working
Recommended
life
compound
Normal
RepliSet-F5
Short
RepliSet-F1/GF1
Normal
Normal
RepliSet-T3
Short
RepliSet-T1/GT1
Normal

Publicité

loading