Description
RepliSet is a specially formulated fast curing two-part silicon
rubber for fl exible high-resolution replicas.
Safety Measures
Storage and Transportation
Store the cartridges horizontally and at room temperature.
Temperature cycling between ambient and a low temperature may
cause air to be drawn into the cartridge. Such cycling should be
minimised.
Lifetime
RepliSet should be used before the expiry date to ensure optimal
product properties. The expiry month and batch code are indicated
on each cartridge.
Expired cartridges may still produce high-quality replicas,
however Struers cannot guarantee for proper curing. If an expired
compound cures normally, the copy (area of interest on the
replica) will normally be OK.
Store opened cartridges with the nozzle on as a seal. An opened
cartridge can be stored at room temperature for at least four
weeks until it is used again.
Precautions
Do not attempt to press material into nozzles, which are blocked
with cured material. Curing agent can be pushed into the
compound causing curing of the material in the cartridge.
Standard
RepliSet is an accepted replicating system for ASTM standard
E 1351.