RepliSet
Users Guide
Gebrauchsanweisung
Guide de l'utilisateur
Dispenser handle
Handgriff
Poignée du distributeur
Operating the 265 ml dispensing system
1.
Push the slide release button and push the slide back
until the piston is in contact with the dispenser body.
2.
Slide the cartridge vertically into position on the front of
the dispenser with the appropriate pistons opposite
each other.
3.
Remove nozzle nut from cartridge (twist counter-
clockwise). Attach mixing nozzle to the cartridge by
aligning notch on nozzle with cartridge lug. Remove
plug from nozzle nut and attach nut. Screw tightly.
4.
Operate dispensing gun until compound issues. When
using a new cartridge discard the first 5 ml issuing from
the nozzle.
5.
Operate the dispenser smoothly to apply the RepliSet
compound. Point the nozzle downwards and keep it
touching or as close to the surface as possible to avoid
trapping air in the replica and to force the material into
surface features. Overlap runs to cover larger areas.
Weave the nozzle end from side to side if a wider bead
NOTE
The process time for RepliSet varies with the temperature. Please see detailed information in Instruction Manual.
For correct results do not go beyond the actual working time and do not go under the actual curing time.
For more detailed information on the RepliSet system, please refer to the RepliSet Reference Guide www.struers.com/manuals.
Hebel zur Verriegelung der Kolbenstange
Levier de relâche de la glissière
is required but each weave should overlap the previous
one. At vertical surfaces the beads should be placed
from the bottom.
6.
Do not apply excessive hand pressure to the dispenser
as this may damage the operating mechanism.
7.
After use it is advisable to leave the mixing nozzle in
place on the cartridge to act as a seal until the cartridge
is used again. The used nozzle will also extract any
cured material from the cartridge ports as it is removed.
8.
Allow the RepliSet compound to cure (check the
cartridge for curing time), and then carefully peel off the
cured replica from one side. To prevent damage and
loss of the recorded detail, do not touch the replica
surface and store the replica in a plastic bag.
9.
Each time the flow of compound is stopped for a period
longer than the working life a new nozzle must be used.
10. Push slide release lever and pull the piston slide to its
back position. Lift the cartridge.
# 50900018
11.08.2014
Dispenser body
Spender-Körper
Corps du distributeur
Piston slide
Kolbenstange
Glissière du piston
Slide release lever