Why:.This.alarm.has.occurred.because.pump.
has.been.in.pause.mode.for.2.minutes..
To correct the
PAUSE.key.to.silence.alarm.and.place.pump.in.
pause.mode.again..Program.and.use.pump.or.
press.the.ON/OFF.key.to.turn.it.off.
SHUT DOOR (Figure 3-8)
Pump.will.sound.a.dual.tone.beep.repeatedly.and.
display.will.read.SHUT.DOOR..Note:.This.alarm.is.
available.only.on.select.pumps.
Why: This.alarm.has.occurred.because.the.
pump.door.was.opened.while.pump.was.running.
To correct the problem:
PAUSE.key.to.silence.alarm.and.place.pump.
in.pause.mode.then.shut.the.door..Check.door.
is.latched.securely..If.door.will.not.stay.closed,.
check.door.latch.and.latch.pocket.on.pump.are.
not.broken.or.cracked.
Messages
DOSE DONE.(Figure 3-9)
This.message.indicates.a.single.feeding.dose.has.
been.delivered.
To clear pump:.Either.press.the.ON/OFF.key.
to.turn.pump.off,.or.press.the.RUN/PAUSE.key.
to.put.pump.in.pause.mode.before.starting.an-
other.feeding.
NEXT DOSE.(Figure 3-10)
If.the.interval.feeding.feature.is.in.use,.between.feed-
ings.display.will.read.NEXT.DOSE.##.##.hr..(##.##,.
indicates.the.time.in.hours.and.minutes.until.the.next.
dose.begins.)
If.the.currently.programmed.feeding.is.not.desired,.either.press.the.ON/OFF.key.
to.turn.pump.off.or.press.the.RUN/PAUSE.key.to.place.pump.in.pause.mode.
where.adjustments.can.be.made.to.the.settings.
problem:.Press.the.RUN/
Press.the.RUN/
3. ALARMS, MESSAGES, AND INDICATIONS
25
Figure 3-8 SHUT
DOOR Alarm
Figure 3-9 DOSE
DONE Message
Figure 3-10 NEXT
DOSE Message
© Copyright 2013 Moog, Inc.