LOAD SET Alarm
1. Is door closed se-
curely?
NO
2. Is door cracked
or tab on inside of
door broken?
NO FLOW IN or NO FLOW OUT Alarms
4.NO FLOW IN.-.Occlusion.between.delivery.set.and.pump.
4 NO FLOW OUT.-.Occlusion.between.pump.and.patient.
1. Is delivery set
tubing pinched,
kinked, or clogged?
NO
2. Is tab on inside of
door broken?
NO
3. Is pressure sensor
region of cassette
receptacle clean?
YES
4. Are there visible
signs of damage
to pressure sen-
sors area?
NO /
Turn.pump.off..Unload.and.reload.delivery.set.
cassette..Make.sure.door.is.closed.securely..Turn.
pump.back.on.
UNSURE
Contact your healthcare provider or Moog
YES
Contact your healthcare provider or Moog
NO
Check.delivery.set.for.obstructions.or.kinks.in.tubing..
YES
Correct.blockage.and.resume.feeding.
Contact your healthcare provider or
YES
Moog Customer Advocacy at (800) 970-2337
Clean.pressure.sensor.area.with.a.cotton.swab,.soft.
cloth.or.dampened.sponge,.or.wash.entire.pump.under.
NO
running.water...Do.not.use.abrasive.materials.or.harsh.
chemicals.
Contact your healthcare provider or
YES
Moog Customer Advocacy at (800) 970-2337
Disconnect.from.patient.and.fill.a.new.delivery.set.with.food..
Prime.set.and.start.pump..If alarm continues, contact your
NO
healthcare provider or Moog Customer Advocacy at
29
4. TROUBLESHOOTING GUIDE
Customer Advocacy at (800) 970-2337
to receive a new door.
Customer Advocacy at (800) 970-2337
for assistance.
to receive a new door.
to return pump for service.
(800) 970-2337 for assistance.