9. ADDITIONAL TECHNICAL INFORMATION
Guidance and manufacturer's declaration – electromagnetic immunity
The.EnteraLite.Infinity.pump.is.intended.for.use.in.the.electromagnetic.environment.specified.below..The.customer.or.the.
user.of.the.EnteraLite.Infinity.pump.should.assure.that.it.is.used.in.such.an.environment.
Immunity.test
Conducted.RF
3.Vrms
IEC.61000-4-6
150.kHz.to.80.
MHz
Radiated.RF
3.V/m
IEC.61000-4-3
80.MHz.to.2.5.
GHz
Radiated.RF
75.V/m
RTCA/DO-160E.
100.MHz.to.8.GHz
Section.20
NOTE.1.At.80.MHz.and.800.MHz,.the.higher.frequency.range.applies.
NOTE.2.These.guidelines.may.not.apply.in.all.situations..Electromagnetic.propagation.is.affected.by.absorption.and.
reflection.from.structures,.objects.and.people.
. Field.strengths.from.fixed.transmitters,.such.as.base.stations.for.radio.(cellular/cordless).telephones.and.land.mobile.
a
radios,.amateur.radio,.AM.and.FM.radio.broadcast.and.TV.broadcast.cannot.be.predicted.theoretically.with.accuracy..
To.assess.the.electromagnetic.environment.due.to.fixed.RF.transmitters,.an.electromagnetic.site.survey.should.be.
considered..If.the.measured.field.strength.in.the.location.in.which.the.EnteraLite.Infinity.pump.is.used.exceeds.the.
applicable.RF.compliance.level.above,.the.EnteraLite.Infinity.pump.should.be.observed.to.verify.normal.operation..
If.abnormal.performance.is.observed,.additional.measures.may.be.necessary,.such.as.re-orienting.or.relocating.the.
EnteraLite.Infinity.pump.
. Over.the.frequency.range.150.kHz.to.80.MHz,.field.strengths.should.be.less.than.3.V/m.
b
EnteraLite
Infinity
Operator's Manual
®
®
IEC.60601.test.
Compliance.level
level
3.Vrms
3.V/m
No.Equipment.
Category.specified
(75.V/m)
Electromagnetic.environment.-.guidance
Portable.and.mobile.RF.communications.equipment.
should.be.used.no.closer.to.any.part.of.the.EnteraLite.
Infinity.pump,.including.cables,.than.the.recommended.
separation.distance.calculated.from.the.equation.
applicable.to.the.frequency.of.the.transmitter.
Recommended.separation.distance
d=1.2√P
d=1.2√P.
80.MHz.to.800.MHz
d=2.3√P.
800.MHz.to.2.5.GHz
where.P.is.the.maximum.output.power.rating.of.the.
transmitter.in.watts.(W).according.to.the.transmitter.
manufacturer.and.d.is.the.recommended.separation.
distance.in.meters.(m).
Field.strengths.from.fixed.RF.transmitters,.as.
determined.by.an.electromagnetic.site.survey,
be.less.than.the.compliance.level.in.each.frequency.
range.
b
Interference.may.occur.in.the.vicinity.of.equipment.
marked.with.the.following.symbol: