2. DIRECTIONS FOR USE
Step 3:
Turn.bag.upside.down.and.
Tubing.
gently.squeeze..Tilt.bag.as.
Port
needed.to.to.evacuate.air.
through.tubing.port.(Fig-
ure 2-4).
Step 4:
Gently.pinch.teal.colored.
tubing.below." ".symbol..
Hold.this.position.until.air.is.
removed.from.tubing..Gently.
Figure 2-4 Squeeze Bag
Figure 2-5 Pinch Tubing
squeeze.bag.at.same.time.to.
assist.fluid.flow..If.fluid.does.
not.flow,.pinch.pressure.may.
be.too.strong..(Figure 2-5).
NOTE:.Inside.the.teal.col-
ored.tubing,.below.the." ".
symbol.is.the.in-line.occlud-
er..The.in-line.occluder.is.
the.built-in.anti.free-flow.
valve..By.pinching.tubing.
gently,.the.tubing.moves.
Figure 2-6 Tubing
Figure 2-7 Tubing
away.from.the.in-line.oc-
Segment Being Pinched
Segment Being Punctured
cluder.allowing.fluid.to.flow.
by Fingernail
(Figure 2-6)..It.is.important.
to.only.pinch.the.tubing.below.the." ".symbol.to.avoid.damaging.the.in-line.oc-
cluder.
NOTE:.Be.extremely.careful.to.pinch.tubing.using.only.the.flat,.soft.part.of.your.
finger..Do.not.use.fingernails.when.pinching.tubing;.doing.so.may.puncture.the.
delivery.set.(Figure 2-7)..If.you.suspect.that.the.delivery.set.has.been.punctured,.
please.discard.and.use.a.new.set.
NOTE:.Air.may.also.be.removed.from.tubing.using.the.pump.prime.feature..See.
step.7.for.instructions.on.using.the.pump.prime.feature.