3. ALARMS, MESSAGES, AND INDICATORS
NO FOOD.(Figure 3-6)
Pump.will.sound.a.dual.tone.beep.repeatedly.and.
display.will.read.NO.FOOD.
Why:.This.alarm.has.occurred.because.pump.
has.detected.air.in.the.tubing..It.takes.approx-
imately.1.ml.of.air,.which.is.approximately.5.
inches.(12.7.cm).in.length.entering.the.teal.col-
ored.tubing.to.cause.an.alarm.
To correct the
PAUSE.key.to.silence.alarm.and.place.pump.in.
pause.mode,.then.check.the.following:.
. . 1..Check.feeding.bag.for.food..If.it.is.empty,.refill.bag.and.prime.the.set.to.
remove.air.from.the.bag.and.tubing.
. . 2..Check.tubing.for.air.bubbles..If.bag.is.full.but.air.is.present.in.the.tubing,.
disconnect.set.from.patient,.press.and.hold.the.PRIME.key.until.air.bub-
ble.has.moved.to.the.end.of.the.tubing.
. . 3... C heck.cassette.receptacle.for.cleanliness,.es-
pecially.around.the.air.sensor..If.cleaning.is.
necessary,.refer.to CHAPTER 5 - CLEANING.
(Page 31).
. . 4...Check.that.an.EnteraLite.Infinity.disposable.set.
is.loaded.into.pump.properly.and.door.is.closed.
. . 5... C heck.disposable.set.for.worn.tubing..If.it.is.
worn,.replace.with.a.new.set.
NOTE:.Blenderized.or.aggressively.mixed.solutions.
may.have.foam..Small.foam.bubbles.may.collect.in.
the.air.sensor.area.and.must.be.cleared.in.order.to.
avoid.a.NO.FOOD.alarm..Allowing.foamy.solutions.to.
sit.for.10.to.15.minutes.after.mixing.and.prior.to.pour-
ing.into.the.bag.will.reduce.the.amount.of.foam.
PUSH RUN TO FEED.(Figure 3-7)
.Pump.will.sound.a.dual.tone.beep.repeatedly.and.
display.will.read.PUSH.RUN.TO.FEED.
EnteraLite
Infinity
Operator's Manual
®
®
problem:.Press.the.RUN/
24
Figure 3-6 NO FOOD Alarm
Figure 3-7 PUSH RUN
TO FEED Alarm