2. DIRECTIONS FOR USE
Feeding Completion:
When.the.dose.has.been.completely.delivered,.
pump.will.stop.running.and.display.will.read.NEXT.
DOSE.then.give.the.number.of.hours.and.minutes.
until.feeding.will.be.repeated.(Figure 2-24)..Display.
will.show.the.run.symbol.flashing.without.the.arcs.
rotating.around.it..Pump.will.repeat.this.cycle.until.
disposable.set.is.empty.or.until.it.is.stopped.by.user.
NOTE:.When.a.feed.interval.has.been.programmed,.
pump.will.not.alarm.DOSE.DONE..Pump.will.cycle.
continuously.as.long.as.feeding.solution.remains.in.
the.bag..When.empty,.pump.will.alarm.NO.FOOD.or.
NO.FLOW.IN.
To.stop.feeding,.press.the.ON/OFF.key.for.1.5.sec-
Figure 2-24 Time
onds.to.turn.pump.off.or.press.the.RUN/PAUSE.key.
Until Next Feeding
to.put.pump.in.pause.mode..Adjustments.to.settings.
can.be.made.at.this.time.
New Feeding Cycle:
Turn.pump.on..Previous.settings.for.rate,.dose,.and.feed.interval.will.be.saved.
in.memory..If.no.changes.are.required,.verify.each.setting.is.correct.and.restart.
pump.running.
NOTE:.Rate,.dose,.and.feed.interval.settings.can.be.locked.into.place.so.that.
changes.cannot.be.made..See.the.User.Preference.Settings.section.(Page 19).
for.instructions.to.change.this.feature.