No.flow.in; No.flow.out - Moog EnteraLite Infinity Manuel De L'opérateur

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

NO FLOW IN.(Figure 3-4)
Pump.will.sound.a.dual.tone.beep.repeatedly.and.
display.will.read.NO.FLOW.IN.
Why:.This.alarm.has.occurred.because.pump.
has.detected.blockage.in.the.set.between.pump.
and.bag..
To correct the
PAUSE.key.to.silence.alarm.and.place.pump.in.
pause.mode,.then.check.the.following:.
. . 1..Check.tubing.for.kinks.or.formula.clumps..
Correct.blockage.
. . 2..Check.cassette.receptacle.for.cleanliness,.
especially.around.the.pressure.sensors..If.
cleaning.is.necessary,.refer.to.CHAPTER 5 -
CLEANING.(Page 31).
NO FLOW OUT (Figure 3-5)
Pump.will.sound.a.dual.tone.beep.repeatedly.and.
display.will.read.NO.FLOW.OUT.
Why:.This.alarm.has.occurred.because.pump.
has.detected.a.blockage.in.the.set.between.
pump.and.patient..
To correct the
PAUSE.key.to.silence.alarm.and.place.pump.in.
pause.mode,.then.check.the.following:.
. . 1..Check.tubing.for.kinks.or.blockages..Remove.
kink.or.blockage.
. . 2..Check.cassette.receptacle.for.cleanli-
ness,.especially.around.pressure.sensors..If.
cleaning.is.necessary,.refer.to.CHAPTER 5 -
CLEANING.(Page 31).
. . 3..Back.pressure.from.patient.may.also.cause.this.alarm..Discuss.with.physician.
problem:.Press.the.RUN/
problem:.Press.the.RUN/
3. ALARMS, MESSAGES, AND INDICATIONS
23