3. ALARMS, MESSAGES, AND INDICATIONS
Alarms
ER01 - ER99 or ERRA - ERRZ (Figure 3-1)
Pump.will.sound.a.continuous.beep.and.display.will.
read.ER.followed.by.a.two.digit.number.or.ERR.fol-
lowed.by.a.letter.to.indicate.a.self-test.has.failed.
To clear alarm:.Turn.the.pump.off.and.then.
back.on..
WARNING: If.an.error.occurs.(ER01.-.ER99.or.
ERRA.-.ERRZ.will.appear.in.display),.all.settings.
and.volume.counters.should.be.checked.and.veri-
fied.before.starting.a.feeding.
LOAD SET.(Figure 3-2)
Pump.will.sound.a.dual.tone.beep.repeatedly.and.
display.will.read.LOAD.SET.
Why: This.alarm.will.sound.when.pump.attempted.
to.run.with.the.set.improperly.loaded.or.missing..
To correct the
PAUSE.key.to.silence.alarm.and.place.pump.in.
pause.mode,.then.check.the.following:
. . 1..An.EnteraLite.Infinity.disposable.set.is.loaded.
into.pump.properly.and.door.is.closed.
. . 2..Check.cassette.receptacle.for.cleanliness,.es-
pecially.around.pressure.sensors..If.cleaning.is.
necessary,.refer.to.CHAPTER 5 - CLEANING.
(Page 31).
LOW BATT.(Figure 3-3)
When.1.hour.or.less.of.battery.life.is.available,.dis-
play.will.flash.LOW.BATT.every.3.seconds.and.pump.
will.beep.every.2.seconds.to.indicate.battery.is.low.
and.pump.will.soon.stop.running..
To correct the problem:.Plug.in.AC.Adapter/
Charger.to.recharge.battery.