2. DIRECTIONS FOR USE
Step 5:
Loop.silicone.tubing.around.pump.wheel.stretch-
ing.lightly..Seat.cassette.into.pump.(Figure 2-8)..
Close pump door.(Figure 2-9).
NOTE:.If.1200ml.set.is.to.be.used.with.Mini.Back-
pack,.load.set.into.back.section.of.pack.and.thread.
cassette.through.port.in.bottom.corner.of.pack.pri-
or.to.loading.cassette.into.pump..
Figure 2-8 Seat Cassette
Step 6:
Press.and.hold.ON/OFF.key.for.1.5.seconds.to.
turn.pump.on..While.pump.runs.through.a.self.test,.
display.will.light.and.an.audible.alarm.will.sound.as.
pump.displays.the.nine.digit.serial.number.three.
digits.at.a.time.for.one.second.each..The.display.
will.then.show.the.letter.'R'.followed.by.a.number,.
which.is.the.software.revision.
Figure 2-9 Close Pump Door
Next,.all.segments.of.display.will.be.shown.for.2.
seconds..Verify all display segments and sym-
bols are active.
The.self.test.is.complete.and.pump.will.then.dis-
play.last.programmed.rate.and.will.be.in.pause.
mode.
If.any.air.is.still.in.the.tubing,.continue.to.Step.7.to.
use.pump.prime.feature.
Figure 2-10 Priming Indication
Step 7:
Press.and.hold.PRIME.key..Alarm.will.sound.once.and.pump.will.begin.pumping.
at.the.maximum.rate.of.600.ml/hr..Display.will.read."TO.STOP.LET.GO".(Figure
2-10)..Once.all.air.is.removed.from.tubing,.release.key..Pump.will.stop,.display.will.
revert.to.last.programmed.rate,.and.pump.will.be.in.pause.mode.
For pump operation instructions:
4.For.a.Single.Feeding.Example:.go.to.Page 10..
4 For.an.Interval.Feeding.Example:.go.to.Page 13.
9