Guidance and manufacturer's declaration – electromagnetic emissions
The.EnteraLite.Infinity.is.intended.for.use.in.the.electromagnetic.environment.specified.below..The.customer.or.the.user.of.the.
Emissions Test
RF.Emissions.
CISPR.11
RF.Emissions.
CISPR.11
Harmonic.emissions.
IEC.61000-3-2
Voltage.fluctuations/Flicker.
emissions.
IEC.61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The.EnteraLite.Infinity.pump.is.intended.for.use.in.the.electromagnetic.environment.specified.below..The.customer.or.the.user.
of.the.EnteraLite.Infinity.pump.should.assure.that.it.is.used.in.such.an.environment.
Immunity.test
Electrostatic.discharge.
(ESD)
IEC.61000-4-2
Electrostatic.fast.
transient/burst
IEC.61000-4-4
Surge
IEC.61000-4-5
Voltage.dips,.short.
interruptions.and.
voltage.variations.on.
power.supply.input.
lines
IEC.61000-4-11
Power.frequency.
(50/60.Hz).magnetic.
field
IEC.61000-4-8
NOTE.U
.is.the.a.c..mains.voltage.prior.to.application.of.the.test.level.
T
EnteraLite.Infinity.should.assure.that.it.is.used.in.such.an.environment.
Compliance
Group.1
Group.B
Class.A
Complies
IEC.60601.test.level
±.6.kV.contact
±.8.kV.air
±.2.kV.for.power.supply.
lines
±.1.kV.for.input/output.
lines
±.1.kV.line(s).to.line(s)
<5.%.U
.(>95.%.dip.in.U
).
T
T
for.0.5.cycle
40.%.U
.(60.%.dip.in.U
).
T
T
for.5.cycles
70.%.U
.(30.%.dip.in.U
).
T
T
for.25.cycles
<5.%.U
.(>95.%.dip.in.
T
U
).for.5.s
T
3.A/m
9. ADDITIONAL TECHNICAL INFORMATION
Electromagnetic environment - guidance
The.EnteraLite.Infinity.uses.RF.energy.only.for.its.internal.function..
Therefore,.its.RF.emissions.are.very.low.and.are.not.likely.to.cause.
any.interference.in.nearby.electronic.equipment.
The.EnteraLite.Infinity.is.suitable.for.use.in.all.establishments,.
including.domestic.establishments.and.those.directly.connected.to.
the.public.low-voltage.power.supply.network.that.supplies.buildings.
used.for.domestic.purposes.
Compliance.level
±.6.kV.contact
±.8.kV.air
±.2.kV.for.power.
supply.lines
±.1.kV.for.input/output.
lines
±.1.kV.differential.
mode
<5.%.U
.(>95.%.dip.in.
T
).for.0.5.cycle
U
T
40.%.U
.(60.%.dip.in.
T
).for.5.cycles
U
T
70.%.U
.(30.%.dip.in.
T
U
).for.25.cycles
T
<5.%.U
.(>95.%.dip.in.
T
U
).for.5.s
T
Class.2.(3.A/m)
39
Electromagnetic.
environment.-.guidance
Floors.should.be.wood,.concrete.or.
ceramic.tile..If.floors.are.covered.with.
synthetic.material,.the.relative.humidity.
should.be.at.least.30%.
Mains.power.quality.should.be.that.
of.a.typical.commercial.or.hospital.
environment.
Mains.power.quality.should.be.that.
of.a.typical.commercial.or.hospital.
environment.
Mains.power.quality.should.be.that.
of.a.typical.commercial.or.hospital.
environment..The.EnteraLite.Infinity.pump.
allows.continued.operation.during.power.
mains.interruptions.via.the.internal.battery.
Power.frequency.magnetic.fields.should.
be.at.levels.characteristic.of.a.typical.
location.in.the.typical.commercial.or.
hospital.environment.
© Copyright 2013 Moog, Inc.