8. IMPORTANT PEDIATRIC CONSIDERATIONS
EnteraLite.Infinity.can.be.used.on.pediatric.patients.if.the.specifications.of.pump.
meet.delivery.requirements.of.the.patient..The.specifications.of.primary.impor-
tance.are:
4.The.flow.rate.range.of.EnteraLite.Infinity.is.0.1.ml/hr.to.10.ml/hr.in.0.1.ml/hr.in-
crements.and.10.ml/hr.to.600.ml/hr.in.1.ml/hr.increments.
4.The.EnteraLite.Infinity.pump.delivers.the.dose.at.the.specified.rate.within.+/-.
5%.volumetric.flow.rate.accuracy.under.the.following.conditions:
.. 8.. utilizing.Moog.EnteraLite.Infinity.disposable.sets
.. 8.. fluid.head.height.at.+6.0.inches.±.0.3.inches.(+15.24.cm.±.0.76.cm)..
with.respect.to.center.of.rotor.
Note:.Head.height.effect.on.accuracy.is.shown.below.(Figure 8-1).
+2.5%
-2.5%
If.these.specifications.meet.the.required.feed.regimen,.EnteraLite.Infinity.can.be.
used.to.deliver.enteral.formulas.to.pediatric.patients..ALWAYS VERIFY RATE,
DOSE AND FEED INTERVAL BEFORE PROCEEDING TO FEED.
-12.in..
(-30.5.cm)
Normal
Figure 8-1 Head Height Effect on Accuracy
8. IMPORTANT PEDIATRIC CONSIDERATIONS
+12.in..
(+30.5.cm)
37
© Copyright 2013 Moog, Inc.