2. DIRECTIONS FOR USE
Next Feeding:
Turn.pump.on..Previous.settings.for.rate,.dose,.and.feed.interval.will.be.saved.
in.memory..If.no.changes.are.required,.verify.each.setting.is.correct.and.restart.
pump.running.
NOTE:.Rate,.dose,.and.feed.interval.settings.can.be.locked.so.that.changes.can-
not.be.made..See.the.User.Preference.Settings.section.(Page 19).for.instructions.
to.change.this.feature.
Operating the Pump

An Interval Feeding Example

CAUTION: Please refer to pages v-vi for additional usage considerations.
If.you.want.to.deliver.100.ml.of.enteral.feeding.solution.at.a.rate.of.50.ml/hr,.and.
repeat.this.feeding.every.6.hours:
Key Sequence for "An Interval Feeding Example"
50 ml/h
100 ml
6.00 hr
Step 1:
Prepare.the.disposable.set.with.enteral.feeding.solution.as.described.in.the.Prim-
ing.and.Loading.the.Disposable.Set.instructions.(Page 7)..
Step 2:
Hang.feeding.bag.or.container.so.that.the.bottom.of.bag.is.at.or.above.the.level.
of.the.pump.door.(refer.to.page.37)..
OR.-.If.an.EnteraLite.Infinity.pack.is.to.be.used,.load.the.pump.and.feeding.bag.
into.the.proper.compartments,.securing.pump,.bag.and.tubing.with.the.pack's.
straps.(see.Page 32).
Step 3:
Turn.pump.on.by.pressing.the.ON/OFF.key..After.running.the.self.test,.display.will.
show.last.programmed.rate.
13
© Copyright 2013 Moog, Inc.