Step 4:
Press.the.+.or.-.key.to.change.the.rate.to.50.ml/hr..
Hold.down.either.key.to.change.rapidly.(Figure 2-18).
Step 5:
Press.the.RATE/DOSE.key.to.display.dose..Press.
the.+.or.-.key.until.a.dose.of.100.ml.is.displayed.(Fig-
ure 2-19).
Step 6:
Press.the.FEED
setting..Press.the.+.or.-.key.until.6.00.hr.is.displayed.
(Figure 2-20).
NOTE:.The.amount.of.time.to.complete.a.rate.and.
dose.combination.can.be.figured.by.dividing.the.dose.
by.the.rate.(Dose/Rate.=.Time)..The.number.of.hours.
to.be.programmed.as.the.feed.interval.is.the.amount.
of.time.to.deliver.the.dose.plus.the.amount.of.time.
pump.should.pause..Example:.100.ml.divided.by.
50.ml/hr.equals.2.hours.for.feeding.delivery,.plus.a.
pause.of.4.hours.equals.a.feeding.interval.of.6.hours.
(100.ml.÷.50ml/hr.=.2.hours.running.+.4.hours.pause.
=.6.hours.from.feeding.start.time.to.start.time).
Pump.will.calculate.number.of.hours.it.will.take.to.deliver.the.rate.and.dose.com-
bination..When.the.FEED
compatible.with.that.combination..In.this.example.the.first.available.value.will.be.
2.00.hr.
NOTE:.Feed.interval.is.displayed.in.hours.and.minutes,.i.e..5.10.=.5.hours.and.
10.minutes.
NOTE:.Feed.interval.cannot.be.programmed.if.rate.and.dose.combination.re-
quires.more.than.24.hours.to.deliver.
Step 7:
Connect.barbed.adapter.to.patient's.enteral.feeding.tube.
EnteraLite
Infinity
Operator's Manual
®
®
INT.key.to.display.the.feed.interval.
INT.key.is.pressed.the.first.available.value.will.be.
14
Figure 2-18 Program
Rate to 50 ml/hr
Figure 2-19 Program
Dose to 100 ml
Figure 2-20 Program
Feed Interval 6.00 hr