CAUTIONS

4. Federal.law.(U.S.A.).restricts.this.device.to.sale.by.or.on.the.order.of.a.phy-
sician,.registered.dietitian,.registered.nurse,.or.other.licensed.practitioner.
4. Dispose.of.EnteraLite.Infinity.Disposable.Sets.properly,.as.required.by.local.law.
4. If.any.leaks.are.detected.in.the.disposable.set,.stop.pump.operation.and.
dis-connect.set.from.patient.and.replace.with.a.new.disposable.set.
4. Care.should.be.used.when.manually.priming.delivery.set.to.ensure.cassette.
is.not.damaged.by.excess.force.
4. Do.not.overload.carry.packs.with.personal.items..Pump.function.may.be.af-
fected.by.kinked.or.pinched.tubing.and/or.unintentional.button.presses.
4. The.EnteraLite.Infinity.uses.an.optical.air-in-line.sensor.to.detect.air.in.the.
tubing.as.it.passes.through.the.pump.causing.the.pump.to.alarm.should.
feeding.be.inadvertently.interrupted.or.when.feeding.is.complete....It.is.im-
portant.to.keep.this.sensor.free.of.debris.including.spilled.fluids.or.food.that.
may.ad-here.to.the.sensor.and.prevent.proper.detection.of.air.in.the.tubing..
.
The.clinician.and.caregiver.should.be.trained.regarding.the.care,.use,.and.
maintenance.of.the.Infinity.pump.and.accessories,.including.proper.cleaning.
of.the.area.around.the.optical.air-in-line.sensor.
.
Clinicians,.care.providers.and.end.users.should.be.aware.of.the.following.
considerations.whenever.using.or.teaching.others.to.use.the.pump:
•.. The.clinician.will.need.to.determine.whether.the.programmed.dose,.infi-
nite.dose,.or.interval.feeding.setting.is.most.appropriate.for.the.patient...
Regardless.of.the.pump.setting,.whenever.a.feeding.is.delivered,.the.
patient.and.pump.delivery.should.be.monitored.with.special.attention.
given.to.those.patients.who.may.not.be.able.to.inform.the.clinician.or.
care.provider.of.discomfort.or.other.concerns.related.to.the.feeding...
•.. If.a.continuous.feed.is.desired.with.the.infinite.dose.setting,.the.delivery.
should.be.monitored.and.proper.amounts.of.additional.food.should.be.
added.prior.to.food.depletion.in.the.bag.or.bottle.
•.. If.the.infinite.dose.setting.is.used.with.a.planned.discontinuation.after.
food.is.delivered.from.the.bag.or.bottle,.allow.all.food.to.move.through.
the.pump...The.pump.will.alarm.when.the.feeding.cycle.is.complete...