2. DIRECTIONS FOR USE
When.you.would.like.to.restart.pump,.press.the.ON/OFF.and/or.the.RUN/PAUSE
key..Pump.will.save.the.memory.of.where.it.was.in.the.feeding.before.stop-
ping..Press.the.RUN/PAUSE.key.and.pump.will.restart.at.the.point.where.it.was.
stopped..
NOTE:.Any.changes.to.pump.settings.during.a.feeding.cycle.will.cause.pump.to.
start.a.new.feeding;.it.will.not.start.where.it.left.off.
Feeding Completion:
When.the.dose.has.been.completely.delivered,.pump.will.stop.running,..and.dis-
play.will.read.DOSE.DONE..Pump.will.be.in.a.pause.mode.until.it.is.turned.off.
or.feeding.is.restarted..To.clear.DOSE.DONE,.press.and.hold.the.ON/OFF.key.
for.1.5.seconds.to.turn.pump.off,.or.press.the.RUN/PAUSE.key.to.put.pump.
in.pause.mode..Adjustments.to.settings.can.be.made.at.this.time,.or.press.the.
RUN/PAUSE.key.a.second.time.to.start.another.feeding.using.same.rate,.dose,.
and.feed.interval.settings.
If.you.would.prefer.the.alarm.to.sound.when.the.dose.
is.complete,.you.can.change.the.pump.dose.com-
plete.alarm.setting.to.BEEP.WHEN.DONE..See.the.
User.Preference.Settings.section.(Page 19).for.in-
structions.to.change.this.setting..When.set.to.BEEP.
WHEN.DONE,.the.pump.will.stop.at.the.end.of.the.
feeding,.beep.intermittently,.and.display.will.read.
DOSE.DONE.(Figure 2-17)..To.silence.alarm.when.
the.dose.is.complete,.press.and.hold.the.ON/OFF.
key.for.1.5.seconds.to.turn.pump.off,.or.press.the.
Figure 2-17 DOSE DONE
RUN/PAUSE.key.to.put.pump.in.pause.mode.
NOTE:.To.program.an.infinite.dose,.press.and.hold.the.+.key.until.display.reads.
INF..When.an.infinite.dose.has.been.programmed,.pump.will.not.alarm.DOSE.
DONE..Pump.will.run.continuously.as.long.as.feeding.solution.remains.in.the.
bag..When.empty,.pump.will.alarm.NO.FOOD.or.NO.FLOW.IN.
12
EnteraLite
Infinity