3. ALARMS, MESSAGES, AND INDICATIONS

Indications

CHEK.(Figure 3-11)
Pump.will.sound.a.dual.tone.beep.once.and.CHEK,.
RATE,.DOSE,.and.FEED.INTERVAL.will.all.blink.on.
the.display.
Why:.This.alarm.sounds.when.the.RATE,.
DOSE,.and.FEED.INTERVAL.are.not.com-
patible..As.discussed.in.the.Interval.Feeding.
Example.(Page.13),.pump.divides.the.dose.
by.the.rate.to.determine.the.amount.of.time.required.for.the.dose.to.be.
delivered..Based.on.the.RATE.and.DOSE.combination,.when.the.FEED.IN-
TERVAL.is.being.programmed,.the.first.value.available.will.be.the.first.value.
that.is.compatible.with.the.RATE.and.DOSE.settings..Should.the.RATE.or.
DOSE.be.changed.after.the.FEED.INTERVAL.has.been.programmed,.and.
the.combination.requires.the.delivery.time.to.be.longer.than.the.amount.of.
time.programmed.as.the.FEED.INTERVAL,.user.will.receive.this.alarm.
For.example,.if.rate.is.100.ml/hr,.and.dose.
is.200.ml,.then.the.pump.will.take.2.hours.to.
deliver.the.dose..The.first.available.FEED.IN-
TERVAL.will.be.2.00.hr..However,.if.the.dose.
is.changed.to.300.ml,.the.dose.will.now.take.3.
hours.to.deliver.and.therefore.2.00.hr.is.an.im-
possible.FEED.INTERVAL..If.the.RUN/PAUSE.
key.is.pressed.before.the.FEED.INTERVAL.has.
been.changed.(the.next.available.value.being.
3.00),.the.pump.will.give.the.CHEK.alarm.
To correct the problem:.Press.the.RUN/
PAUSE.key.to.put.pump.in.pause.mode,.and.re-
program.the.settings.
TO STOP LET GO.(Figure 3-12)
When.pressing.and.holding.the.PRIME.key,.alarm.will.
sound.once,.pump.will.begin.pumping.at.maximum.rate.
of.600.ml/hr.and.display.will.read."TO.STOP.LET.GO"..
To stop priming:.Release.the.PRIME.key.
EnteraLite
Infinity
Operator's Manual
®
®
26
Figure 3-11 CHEK
Indication
Figure 3-12 Priming
Indication