Intrinsically safe and suitable for use in Class1, Groups A-D; Class II, Groups E-G; Class III;
Divisions 1 and 2, Zones 1 & 2 in accordance with entity requirements and METTLER TOLEDO
Control Drawing #72191600.
Refer to the last section of this chapter for the approval entity values associated with the IND560x
terminal.
2.3.1.
Special Conditions (FMus)
•
•
2.4.
European Approvals
Approval details for the IND560x terminal when installed according to European requirements are
listed in this section. The European safety approvals for the IND560x are based upon entity values
and fiber optic power restriction. The IND560x terminal has been examined by the European
notified body DEKRA EXAM GmbH and has been certified as:
Refer to the last section of this chapter for the approval entity values associated with the IND560x
terminal.
2.4.1.
Special Conditions (ATEX)
The following special conditions must be followed when the IND560x is being installed inside a
hazardous area:
•
•
•
•
•
316061929
|
12
|
01/2019
AEx ib IIC T4
•
IS / I, II, III / 1 / ABCDEFG / T4 Ta = 40˚C - 72191600; Entity
•
Avoid electrostatic charging during operation and maintenance.
Operation is only permitted when operational and process related electrostatic charges are not
present
II 2 G Ex ib [op is Ga] IIC T4 Gb
•
II 2 D Ex ib [op is Da] IIIC T60°C Db
IP65
Certificate BVS 09 ATEX E 010X and supplements
•
Avoid electrostatic charging during operation and maintenance.
Operation is only permitted when operational and process related intensive electrostatic charges
are eliminated.
The terminal type IND560x-PANEL has to be mounted into a cabinet/enclosure which is
certified for use as Category II 2D. Mounting must be completed with the gasket included with
the IND560x panel mount. After mounting, IP65 protection must be guaranteed.
UV light must be avoided.
Only cable glands and blanks shall be used which are certified for that purpose.
METTLER TOLEDO IND560x Terminal Installation Manual
2-3