Reference Equivalent Threshold Values For Transducers; Maximum Hearing Level Settings Provided At Each Test Frequency - Interacoustics AS608 Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 199

Reference Equivalent Threshold Values for transducers

Calibration
TDH39
standard
ISO 389-1: 1998
Coupler
IEC60318-3:
standard
1998
Frequency
[dB re. 20 µPa]
[Hz]
125
45.0
250
25.5
500
11.5
750
7.5
1000
7.0
1500
6.5
2000
9.0
3000
10.0
4000
9.5
6000
15.5
8000
13.0

Maximum hearing level settings provided at each test frequency

Frequency
Air Conduction
Hz
TDH39
125
70
250
90
500
100
750
100
1000
100
1500
100
2000
100
3000
100
4000
100
6000
100
8000
90
AS608 - Instructions for Use - EN
TDH39
DD65 v2
ANSI S3.6: 2004
ANSI S3.6 208
ANSI 9A
IEC60318-1
[dB re. 20 µPa]
[dB re. 20 µPa]
45.0
30,5
25.5
11.5
8.0
7.0
6.5
9.0
10.0
9.5
15.5
13.0
Air Conduction
Air Conduction
DD65 v2
DD45
70
90
100
100
100
100
100
100
100
85
70
DD45
PTB Test report
1.61- 4039503/09
IEC60318-3:
1998
[dB re. 20 µPa]
47.5
17
27.0
8
13.0
5,5
6.5
4,5
6.0
2,5
8.0
2,5
8.0
2
8.0
9,5
9.0
21
20.5
21
12.0
Air Conduction
HDA280
70
70
90
90
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
90
90
80
HDA280
Insert EARtone
PTB
EAR3/EAR5
ISO389-2: 1994
IEC60318-3:
60318-5: 2006
1998
[dB re. 20 µPa]
[dB re. 20 µPa]
38.5
26.0
25.0
14.0
13.0
5.5
9.0
2.0
7.5
0.0
9.5
2.0
8.0
3.0
6.5
3.5
9.5
5.5
19.0
2.0
18.0
0.0
Air Conduction
Air Conduction
EARtone 3A
EARtone 5A
70
70
90
90
100
100
100
100
100
100
100
100
100
100
100
100
100
100
95
95
70
80