Cleaning; What To Do If - Seca 954 Mode D'emploi

Masquer les pouces Voir aussi pour 954:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 86

8. CLEANING

9. WHAT TO DO IF ...?

Malfunction
... no weight is displayed
during weighing?
0. 0 0
...
does not appear
before weighing?
... a segment lights up
continuously or not at
all?
...
is displayed?
bAtt
...
is displayed?
St0p
...
is displayed?
7. Confirm your selection each time.
After confirming your selection for Minute you will
leave the menu automatically.
These selections will be automatically transmitted
to the wireless printer.
The wireless printer automatically adds the date
and time to every printout.
NOTE:
For further operation of the wireless printer see
its user manual.
Clean the seat and frame on an as needed basis using
a household cleaner or a commercial disinfectant. Fol-
low the manufacturer's instructions.
Do not use scouring or caustic cleaning products,
spirit, petrol or similar substances for cleaning under
any circumstances. Such products may damage the
high-quality finishes.
Cause/Remedy
The scale has no power supply.
- Check whether the scale is switched on
- Check whether the rechargeable battery
block is inserted
A load was placed on the scales before being
switched on.
- Remove load
- Switch scales off and back on again
There is a fault at that point.
- Notify maintenance service.
The voltage of the rechargeable battery block
drops.
- Charge the rechargeable battery block as
soon as possible
Rechargeable battery block is empty.
- Charge rechargeable battery block
The maximum load has been exceeded.
- Reduce load
English
Cleaning • 79

Publicité

Table des Matières
loading

Table des Matières