Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soapsuds.
WARNING: Do not wash the cuff in a washing machine or
dishwasher!
Accuracy test
We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).
Disposal
Electronic devices must be disposed of in accordance with
the locally applicable regulations, not with domestic waste.
6. Guarantee
This device is covered by a 2 year guarantee from the date of
purchase. During this guarantee period, at our discretion, Microlife
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
Transport costs and risks of transport.
Damage caused by incorrect application or non-compliance
with the instructions for use.
Damage caused by accident or misuse.
Packaging/storage material and instructions for use.
Regular checks and maintenance (calibration).
Accessories and wearing parts: Tubes, pumping ball, cuff.
Should guarantee service be required, please contact the dealer
from where the product was purchased, or your local Microlife
service. You may contact your local Microlife service through our
website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
4
7. Technical Specifications
≤ 540 g
Weight:
Dimensions:
185 x 75 x 110 mm / ± 1 mm
Storage conditions:
-20 - +55 °C
10-80 % relative maximum humidity
Operating conditions: 10 - 40 °C
Measurement range:
0 - 300 mmHg
Resolution:
2 mmHg
Static accuracy:
within ± 3 mmHg between 18 - 33 °C;
within ± 6 mmHg between 34 - 46 °C
Pressure reduction
rate:
2-3 mmHg/sec.
Air leakage:
< ± 4 mmHg/min
Hysteresis error:
within 0 - 4 mmHg
Included accessories: M-cuff (22-32 cm), pump ball, valve,
stethoscope, softbag
Reference to
standards:
EN ISO 81060-1; ANSI / AAMI SP09
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.