Malfunction / Troubleshooting; Safety, Care, Accuracy Test And Disposal; Safety And Protection; Device Care - Microlife BP AG1-20 Mode D'emploi

Table des Matières

Publicité

Note the reading on the manometer 1 as soon as you hear a
faint, rhythmic tapping or thumping sounds. This is the
systolic blood pressure reading.
6. Allow the pressure to continue dropping at the same deflation
rate. Note the reading on the manometer 1 as soon as the
thumping sound stops. This is the diastolic blood pressure
reading.
7. Deflate the cuff completely.
8. Repeat the measurement at least two further times and record
your values, date and time immediately after finishing the
measurements.
9. Remove the cuff and the stethoscope.

4. Malfunction / Troubleshooting

If problems occur when using the device, the following points
should be checked and if necessary, the corresponding measures
are to be taken:
Description
Potential cause and remedy
 Check the ear pieces if they are dirty
The sound transmis-
sion is poor, distorted
or there is extraneous
 Check the tube if it is broken or
noise.
 Check the chest piece if there is any
 Make sure the chest piece is in
 Make sure that the valve is closed.
The pressure does not
 Make sure the cuff is properly
rise although the pump
ball 3 is pumping.
 Check if the cuff, tube and/or bulb is
The deflation rate can
Disassemble the valve from pump ball
not be set to 2-
to check if there is any blockage in the
3 mmHg/sec. by
airway of the valve. Clean the blockage
adjusting the valve 4.
and try again. If it still does not work,
replace it to avoid inaccurate readings.
BP AG1-20
or cracked. If not, make sure you
wear them properly.
twisted.
damage.
contact with the skin and lies on the
brachial artery. Clean or replace any
defective parts if found to avoid inac-
curate reading.
connected to bulb and manometer.
leaking. Replace the defective parts
if any.
Description
Potential cause and remedy
 Make sure that the valve is
The manometer needle
is not at 0 ± 3 mmHg at
 If still more than 3 mmHg deviation,
rest.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.

5. Safety, Care, Accuracy Test and Disposal

Safety and protection

 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section.
 Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
 The cuffs are sensitive and must be handled with care.
 Only pump up the cuff once fitted.
 Never inflate beyond 300 mmHg.
 Always deflate the cuff completely before storage.
 Do not use this device if you think it is damaged or notice
anything unusual.
 Never open this device.
 Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.

Device care

Clean the device only with a soft, dry cloth.
completely open for zero check.
contact your dealer to recalibrate the
manometer.
EN
3

Publicité

Table des Matières
loading

Table des Matières