Operation Mode Display - Burkert 8225 Notice D'utilisation

Transmetteur de conductivite
Masquer les pouces Voir aussi pour 8225:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4 OPERATION

4.2 Operation Mode Display

The following units are indicated within the operation mode display:
12.6 m S
20.6 °C
10.32 m A
HOLD
8225-COMP-**
Conductivity in the required engineering unit. If "0000" or "9999"
is displayed, input value is to low, respectively to high, change
unit or decimal point position
Temperature in °C or °F. If "---- ° C" or "---- ° F" is displayed,
temperature is out of range (-50...+150) or connection with Pt
1000 is interrupted.
The output signal 4...20 mA is proportional to the conductivity
according to the selected measuring range.
HOLD-function. If this function is activated, the transmitter
generates an output current corresponding to the last value before
the option was entered and the relays are blocked in their last
state. This allows the electrode to be cleaned without interruption
of the process. The display in the operation mode will flash and
there is no access to the parameter definition or the test menu,
while the HOLD-function is activated.
ENTER
HOLD
Press Enter for
5 seconds
ENTER
12.6 m S
To disactivate the HOLD function, enter the "HOLD" option again
and select "HOLD NO".
E-15-
CONDUCTIVITY 8225
HOLD YES
HOLD NO

Publicité

Table des Matières
loading

Table des Matières