A.2
Checking the function
Transmitter Functional Check
The purpose of the following procedure is to verify the performance of the signal Trans-
mitter.
Before any tests can be carried out it is vital to check the status of the unit, its batteries
and basic functionality.
To carry out this test the following are required:
• The Transmitters cable Set.
• A fully charged battery pack .
1. Inspection
• Casing
• Cable Set
• Labels
• Battery hatch
• Battery holder
• Battery contacts
Once the general condition of the Transmitter is established the Audio Visual test can be
performed.
The casing should be free of significant damage.
The cable set should be free of damage to the cable's insulation
and clip shrouds.
The terminals should be free of corrosion.
Body labels must be legible and intact. Display label must be
free of damage and tears.
The hatch must lock into place.
All the battery contacts and springs on the holder must be free
of corrosion and the holder in good condition.
The battery contacts must be free of corrosion.