2 3 5
C E F A R T E N S
Cefar TENS is intended for use in the electromagnetic environment specified below. The
customer or the user of Cefar TENS should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Cefar TENS is intended for use in the electromagnetic environment specified below. The
customer or the user of Cefar TENS should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Declaration - electromagnetic emission
Compliance
Group 1
Class B
Class A
Complies
Declaration - electromagnetic immunity
IEC 60601 test
Compliance level
level
±8 kV contact
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±15 kV air
Electromagnetic environment
- guidance
Cefar TENS uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in nearby
electronic equipment.
Cefar TENS is suitable for use
in all establishments, including
domestic establishments and
those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tie.
If floors are covered with
synthetic material, the
relative humidity should be at
least 30 %.