Mounting the Outlet Kit
(accessory)
ABTDR
Stripped
hose
Pipe
10
AbraPol-10
Instruction Manual
Disconnect main power.
Remove the lower front cover plate.
Remove the Recirculation Cooling Unit, if there is one.
Take the entire unit, and place it in position by letting the two
Allen bolts slide into the slots located in the left rear side.
Tighten the Allen bolts.
Connect to
drain hose
Connect to
Recirculation Cooling Unit
Strip the steel spring from approx. 3 cm of the drain hose and
cut. Bend the cut end of the steel spring inwards towards the
centre of the hose. Mount the drain hose onto pipe A and
clamp the stripped section using a hose clamp.
Take another piece of drain hose and check that it is long
enough to reach the Recirculation Cooling Unit. Mount onto
pipe B. (Strip some steel spring from the hose and secure the
hose as described earlier).
Connect to the Recirculation Cooling Unit.
Connect a piece of drain hose from the outlet at the rear (C)
and lead to the drain.
Reposition the Recirculation Cooling Unit.
Make sure that the hoses do not interfere with the belts.
A
B
C
Lead outlet hose
to the drain