Mounting a Recirculation
Cooling Unit (accessory)
0
90
drain angle
Inlet hose
AbraPol-10
Instruction Manual
Disconnect the machine from the mains power supply before connecting or
disconnecting the recirculation pump.
Place the unit where you find it convenient, either inside the
machine or outside, on the left.
The Recirculation Cooling Unit cannot be positioned outside the machine
when an outlet kit (ABTDR) is also used.
Connect the inlet hose from the dosing arm to the pump of the
Recirculation Cooling Unit and tighten the clamp.
(If the cooling unit is placed outside the machine, pass the
hose under the cabinet).
Lead the electrical cable into the electrical panel and connect
to the terminal points.
(If the cooling unit is placed outside, lead through the angle
on the panel).
Slot the 90
drain angle into the pipe in the top of the
0
reservoir lid. (Note: The 45
Recirculation Cooling Unit is not used with this machine).
Cut the drain hose to an appropriate length, and twist it into
the hole located on top of the unit.
(If the cooling unit is placed outside, lead the drain hose
through the hole on the rear left of the machine).
Check that there is a steady fall on the whole course of the
outlet hose.
Check that when the power is turned on, the pump rotates in
the direction indicated by an arrow on the top of the pump. If
not, change two of the phases on the pump motor cable.
Replace the panel/ back plate.
Place a disposable plastic insert in the tank and fold it over
the rim.
Fill the tank with 29.1 l water and 900 ml Struers Additive
for Cooling Fluid.
The water level should be 8-10 cm below the upper rim of the
tank.
Fit the tank lid and sieve.
Replace the front cover.
Too high a level of coolant in the tank might damage the pump. To avoid
this, place the disposable insert so that the pre-punched hole is in front of
the overflow aperture in the tank.
Always maintain the correct concentration of Struers Additive in the cooling
water (percentage stated on the Additive container).
Remember to top up with Struers Additive each time you refill with water.
Warning
Note!
elbow pipe delivered with the
0
IMPORTANT
7