Struers AbraPol-10 Mode D'emploi page 269

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TRANSFORMER INPUT
X1
X1A
F1
CL1
UAC-01
1
STIK 5 POL
2,5AT
24Vac
4AT
GND-S
X1B
CL2
UAC-00
2
STIK 5 POL
GND-S
X1C
3
STIK 5 POL
X1D
F2
CL3
24Vac
4
UAC-1
4AT
STIK 5 POL
4AT
GND-S
X1E
CL4
COMMON
5
UAC-0
STIK 5 POL
GND-S
MOUNTING CONNECTIONS
SKRUEHUL_1
SKRUEHUL_2
SKRUEHUL_3
SKRUEHUL_4
SKRUEHUL_5
SKRUEHUL_6
SKRUEHUL_7
SKRUEHUL_8
GND-S
GROUND CONNECTIONS
R9
1M
GND-P
R10
1M
GND
GND-S
UAC-01; +5V,
1AT
+15V, 300mAT
-15V, 100mAT
+24V, 1,5AT
UAC-1; 24Vac, 4AT (CONTACTOR MAIN, DISC MOTOR,
SAMPLE MOVER , RECIRK. AND WATER-VALVE)
+5VDC, +15VDC AND +24VDC SUPPLY
UAC-01
RV1
C1
-
S10K35
100nF
UAC-00
-15VDC SUPPLY
C11
UAC-01
120uF/50V
V7
V8
UAC-00
1N4007
1N4007
V9
V10
1N4007
1N4007
C14
120uF/50V
HS1
Heatsink
V2
KBU6B
N1
LM2576HV-ADJ
4
FB
+
1
Vin
2
OUT
GND
ON/OFF
3
5
STJ1
GND
+24V
C5
+
120uF/50V
V4
C7
+
2200uF/50V
ICTE-36
+
C8
120uF/50V
R3
10K/1%
+15V
R5
15K/1%
R7
4K75/1%
GND
N4
LM7915
-15V
2
3
IN
OUT
GND
1
C12
C13
470uF/50V
22uF/40V
GND
L2
100uH
R2
28K/1%
+
C2
470uF/50V
V3
SB330
R1
1K5/1%
1
3
IN
OUT
N2
GND
LM7815
2
C6
+
22uF/40V
N3
LM2575HVT-5.0
4
FB
1
IN
2
O
GND
ON
L1
400uH/1A
V5
+
C9
V6
31DQ06
C10
1200uF/16V
100nF
R4
1K/1%
+24V
R6
3,61V
AD-5
page 2
15K/1%
R8
2K1/1%
GND
STRUERS A/S
Rev. A: Baan PDM
VALHOEJS ALLÉ 176
Rev. B: 24V supply changed to switch
DK-2610 ROEDOVRE
mode
DENMARK
PHONE: +45 3670 3500
Rev. C: No change
Rev. D: See page 8
ABRAPOL-10, MACHINE CONTROL BOARD, A2
Rev . E: (FTH 18-08-2004)
POWER SUPPLY
p.6: Namur input circuit redesigned
Size
CAGE Code
FILE NAME.: 5093210.DSN
A2
PAGE1.SCH
Wednesday, August 18, 2004
Scale
+24V
AIR VALVES
DOSING UNIT
GND-P
+15V
GND
VCC
SA-5.0
GND
2,68 - 3,22V
AD-4
page 2
GND
2,95V
AD-6
page 2
DWG NO
Rev
509-3210
E
SLN / SLN
Sheet
1
of
8

Publicité

Table des Matières
loading

Table des Matières