Error Messages
Messages
Errors
Fatal Errors
Message
Process in Progress
Process already stopping
Water level in tank too low
Process is stopping
Process is stopping, await flushing
please
Emergency Stop is active
Database is full !
10 steps is maximum !
2-hand-start released too early
Process is finished
56
AbraPol-10
Instruction Manual
Error messages are divided into three classes:
Messages
Errors
Fatal Errors
Messages are intended to inform the operator of the machine's
progress and advise about minor operational errors.
Errors must be rectified before process can continue.
In case of Fatal Errors, process cannot continue before an
authorised technician has rectified the error. Turn off the unit at
the main switch immediately. Do not attempt to operate unit
before technician has rectified problem.
Explanation
Start is not possible as other
process in progress.
Appears if Stop is pressed
repeatedly.
Water level in recirculation tank is
too low.
Stop has been pressed.
The process is stopping but flushing
removes residues of OP-X.
The emergency button was pressed
and the preparation was
interrupted.
The database storing capacity has
been reached.
For details please see Method
Options.
The maximum of 10 steps has been
reached.
The two start buttons have not been
pressed long enough to start the
process.
The process is finished.
Action
Refill circulation water if needed.
After waiting a moment you can
continue.
Pull the button outwards to release
it and follow the instructions in the
pop-up text.
Delete one or more of the methods
– this will free space to store new
methods.
Nb! it is not possible to delete
Struers Methods.
Two or more methods can be used
alternately.
Keep the start buttons pressed until
the sample motor starts.
You can now start another process.