EasyClean 200e
Adjustable parameters
EasyClean 200e
EasyClean 200e
30 404 676 F
4.5.2 SW parameter description
Factory settings for a general installation. To optimize running time, cleaning agent
consumption, and cleaning effect, adjust the variables accordingly.
Designation
Var.
HOLD extension
T1
Rinse time
T2
Residence time
T3
cleaning agent
Transport time
C3
cleaning agent
1
Transport time
C5
Intermediate cycle
2
(alternating air
and water)
Description
Delay time until transmitter recommen-
ces with transmission of values after
a program/cycle has been completed.
This delay ensures that the electrode
signal, and therefore the first measure-
ment value is stable after hold function
has been deactivated.
Duration of rinsing out of the flushing
chamber with water. Can be activated
by pressing the function "water" and
will be automatically activated before
"Clean". Ensures thorough rinsing of
the flushing chamber.
Residence/action time for cleaning
agent in order to obtain an optimal
cleaning result.
Transport time for cleaning agent. This
parameter depends on the tube length
from the valve unit to the housing as
well as on the housing itself (capacity
of the flushing chamber).
Automatic intermediate rinse/purge air
cycle to rinse out the flushing chamber,
and before moving to service position.
Depends on tube length.
31
Default Range
20 sec
0...2 min
1 min
0...2 min
1 min
0...5 min
30 sec. 10...60 sec.
4 sec.
4...8 sec.