Transmitter connection
Transmitter M300 G2 Process
NOTICE
Bridge terminal 10 and 13 in the EasyClean 200e.
Set relay #2 for example on the transmitter to rinse
contact.
Set rinse time to 5 sec. at the transmitter M300
and define the rinse interval according to the process
requirements. Furthermore define rinse time (T2)
and residence time for cleaning agent (T3) on the
EasyClean 200e.
Transmitter M400 G2
NOTICE
Bridge terminal 10 and 13 in the EasyClean 200e.
Set relay #1 for example on the transmitter to rinse
contact.
Set rinse time to 5 sec. at the transmitter M400 G2
and define the rinse interval according to the process
requirements. Furthermore define rinse time (T2)
and residence time for cleaning agent (T3) on the
EasyClean 200e.
Transmitter M400 2-wire
Set OC1 and Digital input 1 for example on the M400
2-wire HART transmitter to rinse contact.
Set rinse time to 5 sec. at the transmitter and define
the rinse interval according to the process require-
ments. Furthermore define rinse time (T2)
and residence time for cleaning agent (T3) on the
EasyClean 200e.
Transmitter M800
NOTICE
1. M800 1-channel Mix version, please change
TB6 to TB5.
2. The M800 Profinet and Ethernet/IP don't support
Cleaning function.
Set relay #2 and Digital input 1 for example on the
M800 2/4-ch transmitter to rinse contact.
Set rinse time to 5 sec. at the transmitter and define
the rinse interval according to the process require-
ments. Furthermore define rinse time (T2)
and residence time for cleaning agent (T3) on the
EasyClean 200e.
EasyClean 200e
M300 G2 Process
TB1
TB2
EasyClean 200e
M400 G2
TB1
TB2
EasyClean 200e
M400 2-wire
TB1
TB1
EasyClean 200e
M800
TB1
TB6