34
Description of keys
02/2021 © METTLER TOLEDO
Printed in Switzerland
Key
Function/process
"In/Out"
Key to switch from the service
to the measuring position.
Press "In/Out"
Reset or interrupt cycles
and "WATER"
in progress.
simultaneously
Transmitter goes into
"HOLD" mode.
Brief rinsing with water and
air (25 sec.).
End of process.
"WATER"
The housing moves into
the service position.
Transmitter goes into
"HOLD" mode.
Rinsing.
End of process.
"Clean"
The housing moves into
the service position.
Transmitter goes into
"HOLD" mode.
Rinsing.
Cleaning.
Rinsing.
End of process.
Message/display on
Comments
the prog. control
display
Run (Measure)
When the housing moves into
the measure position, the
or
"HOLD" contact of the trans-
Service
mitter is lifted after 12 se-
conds. The sensor is rinsed
with air and water (25 sec.).
Rinsing H20
The housing remains in
the service position.
Run or Service
Depending on housing
position.
Rinsing H20
Rinsing H20
Cleaning
Rinsing H20
EasyClean 200e
EasyClean 200e
30 404 676 F