POWER SOURCE
PUMP LIGHT
INDICATOR
If there are still questions or issues please call Ameda's
certified customer service ParentCare team at
(1-866-99-AMEDA).
22
AC POWER ADAPTER
Finesse™ is optimized for use with the AC Power Adapter.
Make sure the AC power adapter is plugged all the way
into the side of your pump base. Make sure there are no
batteries installed in the pump base. Make sure you rotate
the suction dial up to turn on the pump. Try your pump in
a different electrical outlet.
AA BATTERIES
Be sure the batteries have been inserted correctly
as shown inside the breast pump base's battery
compartment. Use new premium batteries (such as
Energizer, Duracell.) Do not use batteries with AC power
adapter plugged into side of pump base. If problem
persists, you can test that the batteries work in another
battery powered appliance.
LOW POWER: The light blinks slowly and the pump
stops working. Turn the Suction dial to the off position.
If using batteries, replace batteries (see section 7.1) or
remove all batteries and use AC power adapter. If using
AC power adapter and a slow blink occurs, check it
according to Troubleshooting Power Source instructions.
HIGH POWER: The light blinks quickly and the pump
stops working. Turn the Suction dial to the off position.
Check AC power adapter according to Troubleshooting
Power Source instructions, make sure to unplug and re-
connect the AC power adapter into the side of the pump
and check all connections are correct. If using batteries,
check them according to Troubleshooting Power Source
instructions.
ERROR: The light changes to a double fast blink and
pump stops working. Turn the Suction dial to the off
position. Unplug and re-connect the AC power adapter
into the side of the pump and check all connections are
correct, then return to operation.
1-866-992-6332