STEL is automatically reset to zero when the instrument is
turned ON.
STEL value is calculated over a 15-minute exposure.
The following formula is used to calculate the STEL value:
(Minute #1 PPM value) + (Minute #2 PPM value) +....
(minute #15 PPM value) / (15 minutes) = PPM STEL value
If the instrument is ON for less than 15 minutes:
the balance of the minute PPM values is set to zero and
the total is divided by 15 minutes.
11) Time Weighted Average, TWA
The TWA reading displays what the instrument has calculated
since power-ON.
• When this page displays, press the TEST button to clear.
"CLR" displays
• TWA value is automatically reset to zero when the instrument is
turned ON.
• TWA value is calculated over an eight-hour exposure
• The following formula is used to calculate the TWA value:
Sum of 1-minute gas readings/ 480 minutes (8 hours) = TWA value
• If the instrument is ON for less than 8 hours (480 minutes), the
balance of the minute ppm values is set to zero,
12) IR mode.
When instrument displays "IR?", press button to enter IR mode.
If IR communications are not detected for three minutes or the
TEST button is pressed, the instrument will exit this mode.
See FIGURE 2-5 for additional details.

Turning OFF the ALTAIR Pro Detector

1. Press and hold the TEST button for three seconds.
"OFF" and the hourglass display.
2. Continue to press the TEST button for two more seconds.
The instrument turns OFF.
NOTE: The Steel Mill version can't be turned OFF without
removing the battery.