Technical Information
Profibus Option
Offset
Size
Unit Res
Byte
Byte
5
1
-
6
1
-
290/306
Range
Description
-
Error Code (bit field)
B
– Too High Load
7
(the gas load is too high for the pump)
B
– Short Circuit
6
(a short circuit happened between the two of the
three motor phases)
B
– Over Voltage
5
(too high bus voltage during braking)
B
– Verify Line Select
4
Not used
B
– Power Fail
3
(internal circuitry failure)
B
– Controller Over-Temperature
2
( "Controller Power Section Temperature" or
"Controller CPU Section Temperature" are over
the limit)
B
- Pump Over-Temperature
1
("Pump Bearing Temperature" or "Pump Body
Temperature" are over the limit)
B
– Check Connection To Pump
0
(the connection between controller and pump
missing)
-
B
– Set Point R1 Status
0
0 = open
1 = closed
B
– Set Point R2 Status
1
0 = open
1 = closed
B
– Vent Valve Status
2
0 = closed
1 = open
Turbo-V 81-AG Rack Controller User Manual / 87-900-986-01