Output Buffer
Tab. 19
Offset
Size
Unit Res
Byte
Byte
0
1
Turbo-V 81-AG Rack Controller User Manual / 87-900-986-01
Range
Description
B7 = Start/Stop
0 = Stop, 1 = Start
B6 = Low Speed
This bit set the pump target speed to "low speed
value" (Output Buffer, Offset 3), otherwise the
target speed is set to "high speed value" (Output
Buffer, Offset 1).
0 =Off, 1 =On
B5 – Soft Start
This bit activates the soft start function which run-
up the pump smoothly. Please activate this
function only if the pump has been stopped for at
least3 months.
0 = On, 1= Off
B4 = Vent Valve Command
0 = Close
1 = Open
Valid only if in Par Buff, Offset-19, B0=1
B3 = Purge Valve Command
0 = Close
1 = Open
Valid only if Par Buff, Offset-19, B0=1
B2 – Not used
Technical Information
Profibus Option
(I)
17
WIN
Serial
000
001
100
122
145
287/306