Technical Information
Controller Connection
P1 - Input
Figure 9
All the logic input to the controller must be connected at J1 mating
connector. With the provided P1 mating connector (shipped with pin
3 and pin 8 shorted) make the connections with AWG 24, (0.24 mm
or smaller wire to the pins indicated in the figure to obtain the
desired capability.
The following table describes the signals available on the connector.

Tab. 3

Pins
1-com
2-com
216/306
P1 Input Connector
Name
Description
Start/Stop
Contact closed = Start
Contact open = Stop
Low Speed
Contact closed = Low
Speed
Contact open = High
Speed
Turbo-V 81-AG Rack Controller User Manual / 87-900-986-01
Electrical
Note
Optically isolated
Active
Vlow, max = 10v
only in
Imax = 5mA
Remote
Mode
Imin (on) = 2mA
operation
Optically isolated
Active
Vlow, max = 10v
only in
Remote
Imax = 5mA
Mode
operation
2
)
(I)