Front Panel Description
Figure 2
1
Keyboard push-button to recall on the display the cycle number, cycle time and pump life.
2
Keyboard push-button for LOW SPEED mode selection. It is active only when the front
panel operation has been selected. Pressed once, the pump runs at "stand-by" speed. To
unselect the mode, press the push-button again.
3
Keyboard push-button to recall on the display the pump current, pump temperature, pump
power and rotational speed. It is always active regardless of the operating mode selected.
Push-buttons 3 and 1, if pressed together for at least 2 seconds, put the controller in a
routine where it is possible to program some operation parameters.
4
Keyboard push-button for START, STOP/RESET mode selection. It is active only when the
front panel operation has been selected. By pressing once the starting phase begins; if
pressed again it stops the pump. If the pump has been stopped automatically by a fault, this
push-button must be pressed once to reset the controller and a second time to restart the pump
5
LCD back-lighted alphanumeric display: dot matrix 2 lines x 16 characters.
Turbo-V 81-AG Rack Controller User Manual / 87-900-986-01
5